RESEARCH ARTICLE

In vitro Assessment of Clasps of Cobalt-Chromium and Nickel-titanium Alloys in Removable Prosthesis

Mohammed Zaheer Kola, Deepti Raghav, Prince Kumar, Fawaz Alqahtani, Mohammed S Murayshed, Tushar Vithal Bhagat

Citation Information : Kola MZ, Raghav D, Kumar P, Alqahtani F, Murayshed MS, Bhagat TV. In vitro Assessment of Clasps of Cobalt-Chromium and Nickel-titanium Alloys in Removable Prosthesis. J Contemp Dent Pract 2016; 17 (3):253-257.

DOI: 10.5005/jp-journals-10024-1836

Published Online: 01-08-2016

Copyright Statement:  Copyright © 2016; The Author(s).


Abstract

Background

It has been since a long time that the use of basemetal alloys has significantly increased in the area of prosthetic science. One of the factors contributing for the increasing use of base-metal alloys is the increasing and high cost factor of noble metals. Although numerous materials have been tried since the recent past, Cobalt-Chromium (CoCr) alloys still form the most commonly used ones because of the numerous advantages they offer. Titanium alloys are also significantly being used in this field because of their high resistance strength and high corrosion resistance. Therefore, we aimed to evaluate the flexure strength and fatigue resistance of clasps made up of CoCr alloys and Nickel-titanium (NiTi) alloys.

Materials and methods

We evaluated the retentive forces of CoCr and NiTi clasps two different sizes engaging retentive undercuts of different depths (0.25 and 0.50 mm). Mandibluar second premolar was prepared to frame the prosthesis for missing mandibular first and second molar. A total of 50 clasps were included in the study out of which 30 were NiTi clasps and 30 CoCr clasps. We evaluated the loading of the force at the rate of half Newton(N)/second at the central part of the specimen until fracturing of the specimen occurred and finally measurement of the modulus of elasticity was also done. Statistical analysis was carried out; unpaired “t” test was used for evaluating the level of significance.

Results

The mean load needed for CoCr clasps was 0.7450 and for NiTi clasp was 0.6140 Kgf for producing a deflection of 0.25 mm. As far as flexibility is concerned, more value was seen in NiTi group than CoCr group. For deflecting the clasp up to 0.50 mm, the mean load needed for CoCr clasps was 1.4102 and for NiTi clasp was 0.8260 Kgf. The results were statistically significant. While measuring the flexibility, more value (p < 0.05) was seen in NiTi group than in CoCr group. NiTi alloy clasps had mean flexural strength of 1640 MPa and modulus of elasticity of 32 GPa.

Conclusion

For producing deflection of 0.25 and 0.50 mm, the mean loading force was higher for NiTi alloys. Although being increasingly used these days, NiTi alloys have limited use as compared with CoCr alloys due to high cost of titanium and certain other limitations.

How to cite this article

Kola MZ, Raghav D, Kumar P, Alqahtani F, Murayshed MS, Bhagat TV. In vitro Assessment of Clasps of Cobalt-Chromium and Nickel-titanium Alloys in Removable Prosthesis. J Contemp Dent Pract 2016;17(3):253-257.


PDF Share
  1. Corrosion of cobalt based alloys. In: Davis J, editor. Corrosion. Materials Park, OH: ASM International; 1987. p. 1. 657-700.
  2. The in vitro toxicity of cobalt-chrome molybdenum alloy and its constituent metals. Biomaterials 1986 Jan;7(1):25-29.
  3. Reaction of fibroblasts to various dental casting alloys. J Oral Pathol 1988 Aug;17(7):341-347.
  4. Cobalt-base alloys used in bone surgery. Injury 2000 Dec;31 Suppl 4:18-21.
  5. Biocompatibility of nitinol alloy as an implant material. J Biomed Mater Res 1976 Sep;10(5):695-731.
  6. Anodic polarization behavior of TiNi and Ti-6Al-4V in simulated physiological solutions. J Dent Res 1980 Oct;59(10):1590-1595.
  7. The super-elastic property of the Japanese NiTi alloy wire for use in orthodontics. Am J Orthod Dentofacial Orthop 1986 Jul;90(1):1-10.
  8. Deflection fatigue of cobaltchromium, titanium, and gold alloy cast denture clasp. J Prosthet Dent 1995 Oct;74(4):412-419.
  9. Mccracken's removable partial prosthodontics. 10th ed. Elsevier Science; 1999. p. 258-265.
  10. Fatigue resistance of cast occlusal rests using CoCr And Ag-Pd-Cu-Au alloys. J Prosthet Dent 2003 Sep;90(3):261-269.
  11. Phillips’ science of dental materials. 11th ed. St. Louis: Elsevier; 2003.
  12. Comparison of cast TiNi alloy clasp retention with conventional removable partial denture clasps. J Prosthet Dent 2004 Apr;91(4):374-382.
  13. Fatigue resistance and flexural behavior of acetal resin and chrome cobalt removable partial denture clasp: an in vitro study. Eur J Prosth 2015;3(3):71-76.
  14. The mechanical properties of the cobalt-chromium alloys and their relation to partial denture design. Br Dent J 1965 Nov 2;119(9):389-396.
  15. Stress-relaxation testing. Part IV: Clasp pattern dimensions and their influence on clasp behavior. J Prosthet Dent 1983 Sep;50(3):319-326.
PDF Share
PDF Share

© Jaypee Brothers Medical Publishers (P) LTD.