ORIGINAL RESEARCH |
https://doi.org/10.5005/jp-journals-10024-3492
|
Research on the Role of Surface Treatment of the Metal Surface on the Strength of the Metal–Ceramic Bond
2Department of Prosthodontics, School of Dentistry, National and Kapodistrian University of Athens, Athens, Greece
3Division of Dental Technology, Department of Biomedical Sciences, University of West Attica, Athens, Greece
Corresponding Author: Georgia Asproudi, 21 Mandilara Nikiforou Str., Egaleo, Athens, Greece, e-mail: gasproudi26@hotmail.com
How to cite this article: Asproudi G, Galiatsatos P, Galiatsatos A. Research on the Role of Surface Treatment of the Metal Surface on the Strength of the Metal–ceramic Bond. J Contemp Dent Pract 2023;24(3):188–194.
Source of support: Nil
Conflict of interest: None
ABSTRACT
Aim: The aim of the present study is to investigate the metal–ceramic bond strength as a result of three different surface treatment methods: (1) oxidation, (2) oxidation and sandblasting, and (3) double oxidation on the metal substrate.
Materials and methods: A total of 72 metal substrates were made from two different types of metal–ceramic alloys (n = 36): group I, Ni–Cr and group II, Co–Cr alloys. Each group was further divided and subjected to three different surface treatments (n = 12): (1) oxidation in accordance with the manufacturer’s instructions; (2) oxidation according to the manufacturer’s instructions and then sandblasting with Al2O3, with a grain size of 110 μm, a pressure of 75 psi for 10 sec with a distance of 5 cm and steam cleaning; and (3) double oxidation. The bond strength of the specimens was evaluated with the three-point bending process. The data were recorded, tabulated, and statistically analyzed.
Results: For group I, the materials with oxidation based on the specifications, show mean value of 64.02 Nt. The oxidation and sandblasting materials have mean 55.92 Nt. The double oxidation materials have mean 55.47. For group II, the materials with oxidation based on the specifications, show mean value of 58.46 Nt. The oxidation and sandblasting materials have a mean value of 42.56 Nt. The double oxidation materials have mean 42.96 Nt.
Conclusion: The best method of treatment of the metal substrate is specification oxidation, in terms of the strength of the metal–ceramic bond. Further treatment of the metal substrate reduces the strength of the metal–ceramic bond.
Clinical significance: A prerequisite for clinical success of metal–ceramic prosthetic restorations is the increased strength of the bond between ceramic material and metal substrate. With that in mind, the present research gives important insight into best practices for prosthetic restorations.
Keywords: Co–Cr alloy, Metal–ceramic bond, Metal substrate, Ni–Cr alloy, Surface treatment.
INTRODUCTION
Metal–ceramic prosthetic restorations have been used for several decades in clinical practice with survival rates of up to 90% for the period of 10–15 years, because of their biocompatibility and esthetics, which are deemed superior to the alternatives.1–6 The strength of these restorations is an important factor for their survival in the dynamically changing oral environment (temperature change, pH change, chemical changes), and even under the application of intense and alternating loads. In particular, the forces developed on dental restorations during the application of masticatory loads are complex (tensile forces, shear forces, lateral forces).1–10 Thus, the choice of the appropriate material and the appropriate technique of fabrication of metal–ceramic restoration is one of the main factors that will ensure their protection against fracture.10–15
Also critical to the clinical longevity of these restorations is the achievement of a satisfactory bond between the alloy and the ceramic material. The treatment of the metal framework prior to the porcelain build-up during the fabrication of these restorations has been shown to be a key step that plays an important role in the strength of the metal–ceramic bond.11–18 It is commonly discussed that when metal surfaces are prepared before applying porcelain, the porcelain clings more securely to the metal and the strength of the metal–ceramic bond is increased. Usually, a degassing technique or metal coating is used to condition the metal. Regarding the exact preparation of metals for ceramic bonding, opinions continue to vary, however.1–3
Various methods of treatment of the metal substrate prior to porcelain construction have been proposed from time to time, such as (a) specification-based oxidation, (b) sandblasting, (c) specification-based oxidation and sandblasting, (d) vacuum oxidation, and (e) double oxidation.14–26
The aim of the present study is to investigate the metal–ceramic bond strength as a result of three different surface treatment methods of the metal substrate. These three methods are (1) oxidation according to the manufacturer’s instructions (the metal is subjected to oxidation heat treatment, which includes degassing, outgassing, and preoxidation. This process cleans the metal’s surface of impurities and forms the metal oxide coating); (2) oxidation and sandblasting (cleaning and roughening components’ surfaces using the force of a high-speed sand flow); and (3) double oxidation on the metal substrate (oxidation, sandblasting, steam cleaning and oxidation again).
The null working hypothesis is that the various treatment methods of the metal surface do not affect the strength of the metal–ceramic bond.
MATERIALS AND METHODS
All the research tests were carried out in the Biomaterials Laboratory of the School of Dentistry of the University of Athens. A total of 72 metal–ceramic specimens wherein 36 comprised group I: Ni–Cr alloy (Keralloy N Supreme, Siladent Co, Germany) and group II: 36 of Co–Cr alloy (Wirobond C, Bego Co, Bremen, Germany) were fabricated according to the EN ISO 9693 specification, with specific dimensions of 0.5 ± 0.05 mm thickness, 3 ± 0.1 mm width, and 25 ± 1 mm length.14 After fabrication, an X-ray inspection (gamma radiation intensity; 60 kV and 7.10 mAs) was performed to determine the internal porosity. The metallic porous substrates if confirmed were remanufactured and X-ray retested to meet the required sample size of the study. Table 1 represents the chemical composition of the alloys considered.
Element (ww%) | Ni | Cr | Mo | Si | Co | W | Fe | Ce | C |
---|---|---|---|---|---|---|---|---|---|
Keralloy N supreme alloy | 61.4 | 25.9 | 11.0 | 1.5 | – | – | – | – | |
Wirobond C alloy | – | 26.0 | 6 | 1 | 61 | 5 | 0.5 | 0.5 | Max 0.05 |
Thirty-six specimens of each group were further divided into three groups (n = 12) based on the specific surface treatments before the application of ceramic material in the methodology.
Group A: Corrosion according to the manufacturer’s guide.
Group B: Oxidation according to the manufacturer’s instructions followed by sandblasting with Al2O3, grain size 110 μm, pressure 75 psi for 10 sec at a distance of 5 cm and cleaning with steam.
Group C: Double oxidation, i.e., oxidation, sandblasting with Al2O3, grain size 110 μm, pressure, 75 psi for 10 sec with a distance of 5 cm, steam cleaning and oxidation again in the same cycle as the original.
The ceramic mass was placed on the metal substrates in the dental laboratory, following the manufacturer’s instructions. The coating of the ceramic material (Ex – 3, Noritake Co, Japan) was applied at a thickness of 1.1 ± 0.1 mm, a width of 3 ± 0.1 mm, and a length of 8 ± 0.1 mm, according to the EN ISO 9693 specification.14
The metal–ceramic specimens were then subjected to thermal cycling: 3000 cycles between 500 C ± 1 and 550 C ± 1oC in distilled water. The residence time at each temperature was 10 sec, while the time to transfer the specimens from one water bath to another was 5 sec.
Τhe 72 metal–ceramic specimens (36 of group I and 36 of group II) were subjected to a three-point bending test to determine the strength of the metal–ceramic bond using a mechanical testing apparatus (Tensiometer 10, Monsanto, Akron, Ohio, USA). The test was carried out in the Biomaterials Laboratory of the School of Dentistry of the University of Athens. A force/load was applied to the specimens at the center, with a head displacement rate of 1.5 ± 0.5 mm/min. The failure of the metal–ceramic bond was perceived by a sudden decrease in the stress and strain curve. Thus, the load at which the failure of the metal–ceramic bond was observed was recorded. The strength of the metal–ceramic bond in units of MPa was calculated on the basis of the relationship:
where Ffail is the load at which failure was observed or otherwise the failure force (Nt) and k is a coefficient defined by EN ISO 9693 which ranges from 0.5 ± 0.05 of the metal substrate and on the young’s modulus of the alloy.14
Fracture surfaces were examined using optical microscopy (Leica M-80 DFC 295, Leica Microsystems GmbH, Germany) to determine the mode of failure based on the fracture origin (adhesive fracture, adhesive/cohesive fracture, cohesive fracture). The obtained results were statistically analyzed by analysis of variance (ANOVA) at a 0.05 significance level with the use of SPSS software.
RESULTS
Results of Optical Microscope for Group Ni–Cr Alloy (Group I).
All specimens in all subgroups exhibited cohesive failure. Inspection of the specimens under the optical microscope showed the persistence, on the fractured metal surfaces, of elements. These most likely represent elements of the adhesive and ceramic mass.
Total Results for Group I (Ni–Cr Alloy).
A scoreboard showing the overall results of Group I can be found in Figure 1.
Statistically significant differences are observed in group I by distinguishing the results according to the three treatment methods applied to the materials.
The materials with oxidation based on the specifications, show mean 64.02 Nt.
The oxidation and sandblasting materials have mean 55.92 Nt. The double oxidation materials have mean 55.47 Nt.
It was noticed that between all three methods there are no significant differences in the mean value but in the range of bending values. Similarly, the specification-based oxidation shows significant differences in addition to the other two processing methods.
The oxidation and sandblasting method, despite its low range of values, seems to have the highest concentration and, despite the limited range, enables the craftsman to have more concentrated strength values of the material.
Similarly, specification-based oxidation gives a wider range of values which means that it is more difficult for the craftsman to determine the range of values that the material can withstand and accordingly to advise the customer on the care he should take with the pressures he applies to the material. Similarly, double oxidation shows the lowest strength values of the other two groups.
It was concluded that in group I, specification-based oxidation seems to have higher strength rates than oxidation and sandblasting and double oxidation. It is obvious that further treatment of the materials reduces the strength of the materials, especially oxidation and sandblasting, while the effect of double oxidation is lower on the strength of the material.
It is important to notice that despite the differences between the three types of treatment on the strength of the material, the average strength value has significant variations, ranging from 64 Nt (288 MPa) to 54 Nt (216 MPa) approximately, while there is a correlation of the average value in subgroups IIB and IC.
Results of Optical Microscope for Group Co–Cr Alloy (Group II)
All specimens from all subgroups exhibited cohesive failure. Inspection of the specimens under the optical microscope showed the persistence, on the fractured metal surfaces, of elements likely to represent elements of opacity and ceramic mass.
Total Results for Group II (Co–Cr)
A scoreboard showing the overall results of group II can be found in Figure 2.
Significant differences are observed in group II by distinguishing the results according to the three treatment methods applied to the materials.
The materials with oxidation based on the specifications, show mean value of 58.46.
The oxidation and sandblasting materials have a mean value of 42.56 Nt.
The double oxidation materials have mean 42.96 Nt.
It was noticed that between all three methods there are no significant differences in the mean value but in the range of bending values. Similarly, the specification-based oxidation shows significant differences with the other two processing methods.
The double oxidation method, despite the low range of values it shows, seems to have the highest concentration and despite the limited range, it enables the artisan to have more concentrated material strength values; also, the double oxidation method shows the lowest material strength values.
Similarly, specification-based oxidation gives a wider range of values which means that it is more difficult for the craftsman to determine more limited range of Nt values that the material can withstand and accordingly advise the customer on the care to be taken in the stresses applied to the material.
It was concluded that in group II, specification-based oxidation seems to have higher strength rates, compared with oxidation and sandblasting and double oxidation. It can be seen that further treatment of the materials reduces the strength of the materials, especially double oxidation, while the effect of oxidation and sandblasting is lower on the strength of the material. Importantly, despite the differences that the three types of treatment show on the strength of the material, the average strength value has significant variations, ranging from approximately 42 Nt (155.4 MPa) to 58 Nt (214.6 MPa), and there is a correlation of the average value in groups IIB and IIC.
Certainly, the occurrence of outliers affects the sample averages, but it is clear that specification-based oxidation enhances the strength of the material more than any other method.
Comparison of the Means of the Two Groups Using the Independent Samples t-Test
At the same time, the purpose was to compare the means of the two independent groups of measurements at hand, to determine whether the means of the populations from which the samples came are significantly different (independent samples t-test).
Group I (Ni–Cr Alloy)
Independent samples t-test criterion was used for groups IΑ (1-1 to 1-12) and 1Β (1-13 to 1-24). The results showed significant differences in resistance values between these groups. Group IA shows mean = 64.02 Nt.
Independent sample t-test criterion was used for groups IΒ and IC. The results showed non-significant differences in resistance values between these groups. Group IB shows mean = 55.92 Nt. No significant difference is shown as t = 0.840 (p) > 0 0.05 (Tables 2–5).
Processing method | N | Mean | Std. deviation | Std. error mean | |
---|---|---|---|---|---|
Force (Nt) | Oxidation and sandblasting | 12 | 64.0283 | 8.46854 | 2.44466 |
Double oxidation | 12 | 55.9233 | 5.30201 | 1.53056 |
Levene’s test for equality of variances | t-test for equality of means | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
F | Sig. | t | df | Sig. (2-tailed) | Mean difference | Std. error difference | 95% Confidence interval of the difference | |||
Lower | Upper | |||||||||
Force (Nt) | Equal variances assumed | 4.083 | 0.056 | 2.810 | 22 | 0.010 | 8.10500 | 2.88426 | 2.12341 | 14.08659 |
Equal variances not assumed | 2.810 | 18.475 | 0.011 | 8.10500 | 2.88426 | 2.05655 | 14.15345 |
Processing method | N | Mean | Std. deviation | Std. error mean | |
---|---|---|---|---|---|
Force (Nt) | Oxidation and sandblasting | 12 | 55.9233 | 5.30201 | 1.53056 |
Double oxidation | 12 | 55.4758 | 5.43613 | 1.56928 |
Levene’s test for equality of variances | t-test for equality of means | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
F | Sig. | t | df | Sig. (2-tailed) | Mean difference | Std. error difference | 95% Confidence interval of the difference | |||
Lower | Upper | |||||||||
Force (Nt) | Equal variances assumed | 0.569 | 0.459 | 0.204 | 22 | 0.840 | 0.44750 | 2.19208 | –4.09861 | 4.99361 |
Equal variances not assumed | 0.204 | 21.986 | 0.840 | 0.44750 | 2.19208 | –4.09877 | 4.99377 |
Independent samples t-test criterion was used for groups IΑ and IC. The results showed significant differences in resistance values between these groups. Group IA shows mean = 64.28 Nt. Significant difference is shown as t = 0.008 (p) < 0.05 (Tables 6 and 7).
Processing method | N | Mean | Std. deviation | Std. error mean | |
---|---|---|---|---|---|
Force (Nt) | Oxidation and sandblasting | 12 | 64.0283 | 8.46854 | 2.44466 |
Double oxidation | 12 | 55.4758 | 5.43613 | 1.56928 |
Levene’s test for equality of variances | t-test for equality of means | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
F | Sig. | t | df | Sig. (2-tailed) | Mean difference | Std. error difference | 95% Confidence interval of the difference | |||
Lower | Upper | |||||||||
Force (Nt) | Equal variances assumed | 2.444 | 0.132 | 2.944 | 22 | 0.008 | 8.55250 | 2.90499 | 2.52792 | 14.57708 |
Equal variances not assumed | 2.944 | 18.750 | 0.008 | 8.55250 | 2.90499 | 2.46678 | 14.63822 |
Group II (Co–Cr Alloy)
Independent sample t-test criterion was used for groups IIΑ (2-1 to 2-12) and 2Β (2-13 to 2-24). The results showed statistically non-significant differences in resistance values between these groups. Group IIA shows mean = 58.46 Nt. It shows no statistically significant difference as t = 0.180 (p) > 0.05 (Tables 8–10).
Processing type | N | Mean | Std. deviation | Std. error mean | |
---|---|---|---|---|---|
Force (Nt) | Oxidation based on the specifications | 12 | 58.4633 | 33.37539 | 9.63465 |
Oxidation and sandblasting | 12 | 42.5467 | 21.79189 | 6.29078 |
Levene’s test for equality of variances | t-test for equality of means | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
F | Sig. | t | df | Sig. (2-tailed) | Mean difference | Std. error difference | 95% Confidence interval of the difference | |||
Lower | Upper | |||||||||
Force (Nt) | Equal variances assumed | 3.890 | 0.061 | 1.383 | 22 | 0.180 | 15.91667 | 11.50653 | –7.94642 | 39.77975 |
Equal variances not assumed | 1.383 | 18.937 | 0.183 | 15.91667 | 11.50653 | –8.17224 | 40.00557 |
Processing method | N | Mean | Std. deviation | Std. error mean | |
---|---|---|---|---|---|
Force (Nt) | Oxidation and sandblasting | 12 | 42.5467 | 21.79189 | 6.29078 |
Double oxidation | 12 | 42.9683 | 28.07332 | 8.10407 |
Group independent sample t-test criterion was used for groups IIB (2-13 to 2-24) and 2C (2-25 to 2-36). The results showed statistically non-significant differences in resistance values between these groups. Group IIB shows mean = 42.54 Nt. It shows no significant difference as t = 0.968 (p) > 0.05 (Tables 11).
Levene’s test for equality of variances | t-test for equality of means | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
F | Sig. | t | df | Sig. (2-tailed) | Mean difference | Std. error difference | 95% Confidence interval of the difference | |||
Lower | Upper | |||||||||
Force (Nt) | Equal variances assumed | 0.646 | 0.430 | –0.041 | 22 | 0.968 | –0.42167 | 10.25913 | –21.69781 | 20.85447 |
Equal variances not assumed | –0.041 | 20.725 | 0.968 | –0.42167 | 10.25913 | –21.77394 | 20.93060 |
Independent samples t-test criterion was used for groups IIΑ (2-1 to 2-12) and 2C (2-25 to 2-36). The results showed non-significant differences in resistance values between these groups. Group IIA shows mean = 58.46. It shows no significant difference as t = 0.231 (p) > 0.05 (Tables 12 and 13).
Processing method | N | Mean | Std. deviation | Std. error mean | |
---|---|---|---|---|---|
Force (Nt) | Oxidation and sandblasting | 12 | 58,4633 | 33,37539 | 9,63465 |
Double oxidation | 12 | 42,9683 | 28,07332 | 8,10407 |
Levene’s test for equality of variances | t-test for equality of means | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
F | Sig. | t | df | Sig. (2-tailed) | Mean difference | Std. error difference | 95% Confidence interval of the difference | |||
Lower | Upper | |||||||||
Force (Nt) | Equal variances assumed | 1.068 | 0.313 | 1.231 | 22 | 0.231 | 15.49500 | 12.58977 | –10.61459 | 41.60459 |
Equal variances not assumed | 1.231 | 21.373 | 0.232 | 15.49500 | 12.58977 | –10.65908 | 41.64908 |
DISCUSSION
In the present study, the effect of surface treatment of the metal frame on bond strength in metal–ceramic restorations was investigated. The present study was done using two common categories of alloys for metal–ceramic restorations (Ni–Cr and Co–Cr) and ceramic material (Noritake Ex-3 porcelain), through the three-point bending test method.
In the literature, there are several laboratory tests to check the strength of the metal–ceramic bond. The most used method is the one proposed by the international standardization organization EN ISO 9693 and is the three-point bending method, where the developing stresses can be broken down into shear and tensile ones.14 The reason why this method is considered more appropriate among laboratory methods is because it is more similar than the others, in terms of the development of stresses on the porcelain–alloy interface. Additionally, the three-point in vitro bending test is a clear indication of the metal–ceramic bond strength for the behavior of structures clinically.7,17,20
In the three-point bending test, the part of the test piece to which the force (metallic) is applied and compressed, unlike the ceramic coating, is in a tensile state, while the edges of the test piece are in a shear state. From the energy developed in the form of tension through the application of the constant force to the metal surface of the specimen, a part of it is allocated for bending the ceramic material. Another part of it is allocated for the bending of the metal substrate, and finally a small percentage of the force is lost in the form of frictions.7,17,20
On basis of the foregoing, the bending strengths of the ceramic material (breakage) and the metal substrate (deformation) must be greater than the strength of their bonding in order for the system to be evaluated. Second, the binding strength of the two different materials must exceed the minimum acceptable limit of 25 MPa to be successful (EN ISO 9693). This was taken into account and confirmed during the manufacture of metal substrates in this study.
Specifically, for both categories of alloys, the ANOVA test was used to draw conclusions about whether the method of oxidation of the metal substrate (specification-based oxidation – oxidation and sandblasting – double oxidation) affects the strength of the metal–ceramic bond. Dispersion analysis with one factor showed that in both groups of alloys (Ni–Cr and Co–Cr) the method of oxidation based on specifications has the highest rates of resistance compared with the method of oxidation and sandblasting and the method of double oxidation. In addition, it was found that there is a uniqueness in the range of values, making the test pieces of both groups, regardless of the oxidation method, durable, since the minimum strength values presented were higher than the minimum acceptable limit of 25 MPa of the EN ISO 9693 specification. The mean strength value of the Co–Cr group ranged from 42 Nt to 58 Nt, while that of the Ni–Cr group from 44 Nt to 54 Nt. Owing to this fact, it was deemed necessary to use the independent samples t-test, which showed a slight difference in terms of the strength of the metal–ceramic bond in the test pieces of both groups, in favor of the oxidation method based on specifications.
This difference in strength of the metal–ceramic bond recorded during the experimental procedure in favor of the group of specimens that the metal substrate oxidized according to the manufacturer’s instructions, may most likely be related to the thickness of the created oxide layer in the alloy and ceramic mesophase.26–30
In all metal–ceramic alloys, oxides develop during the oxidation stage, the thickness of the layer of which plays an important role in the strength of the metal–ceramic bond.16,21,26 The increase in the thickness of the oxide band in the same alloy leads to a decrease in the strength of the metal–ceramic bond, due to reduced transcrystalline actions, from the deformations caused by O2 in the crystal lattice of oxides. But also, the reduction of the thickness of the oxide band results in the creation of very few retaining molecules and consequently reduced strength of the metal–ceramic bond. It is known that chromium (Cr) creates oxides that do not favor the metal–ceramic bond, and to reduce the negative action its added to alloys containing chromium elements with high chemical affinity for oxygen, such as Hf, Yt, Ti, and Be.3,5,6,10,15,16,21,26 The average thickness and strength of the oxide belt play a fundamental role in the strength of the metal–ceramic bond. The percentage of oxides is controlled mainly by regulating the temperature and the presence of oxygen (vacuum formation) during the oxidation stage.
Thus, the oxidation of the alloy in strict compliance with the manufacturer’s instructions leads to a controlled thickness of the oxide layer, which is what is required for a strong metal–ceramic bond. On the contrary, even a small deviation (increase or decrease) of the oxidation temperature can lead to disturbances of the oxide layer and affect the metal–ceramic bond. An additional parameter that can modify the behavior of the oxide layer is sandblasting, which can often contaminate the surface of the cast, developing disturbances in the alloy–porcelain connection mechanism.3,5,16,21,26
Reviewing the literature, it is established that the strength of the metal–ceramic bond with various methods of processing the metal substrate has become the subject of research by many researchers.30–34
With regard to the determination of the failure type, the examination of specimens under the optical microscope showed that all test pieces in all three methods used showed cohesive failures. These elements prevailed throughout the surface, since elements of opacity and ceramic mass were observed. This is confirmed by the images obtained from the observation of the specimens with the optical microscope.
The strength of prosthetic restorations is important for their survival in the dynamically changing environment of the human body (temperature change, pH change, chemical changes), especially under the exercise of alternating loads. In particular, the trends that develop in dental restorations during the exercise of chewing loads, are complex. Thus, the selection of the appropriate artificial construction of dental restorations, as well as the appropriate material to be used for their construction, can ensure their protection from breakage.25–30
An important limitation of the present study is the fact that only three methods were tested and as such future studies should expand their scope. Additionally, retesting is required to verify these results.
CONCLUSIONS
The present study concludes that the strength of the metal–ceramic bond of the metal substrates, regardless of the treatment of the metal surface, covers the lower permissible strength limits of the metal–ceramic bond (25 MPa), according to EN ISO 9693, the type of failure for both groups of alloys is cohesive type, which confirms the strength of the bond. Additionally, among the three examined methods of processing the metal surface oxidation shows a superiority in the strength of the metal–ceramic bond (statistically insignificant) based on specifications, due to an increased value of shear strength. In conclusion, it should be stated that to achieve a satisfactory and strong bond in metal–ceramic prosthetic restorations, it is necessary to strictly follow the manufacturer’s instructions during the oxidation stage of the alloy.
REFERENCES
1. Abbott SJ. Metal-ceramic restorations. J Prosthet Dent 1978 Mar;39(3):293294. DOI: 10.1016/s0022-3913(78)80099-7. PMID: 273091.
2. Reitemeier B, Hänsel K, Kastner C, et al. A prospective 10-year study of metal ceramic single crowns and fixed dental prosthesis retainers in private practice settings. J Prosthet Dent 2013;109(3):14955. DOI: 10.1016/S0022-3913(13)60034-7. PMID: 23522363.
3. Bai R, Sun Q, He Y, et al. Ceramic toughening strategies for biomedical applications. Front Bioeng Biotechnol 2022;10:840372. DOI: 10.3389/fbioe.2022.840372.
4. Goiato MC, Alves Pesqueira AA, Monteiro DR, et al. Clinical satisfaction and quality of ceramic fixed dentures. Int J Appl Ceram Technol 2014, 11:100105. DOI: 10.1111/ijac.12019.
5. Fisher H, Dautzenberg G, Marx R. Nondestructive estimation of the strength of dental ceramics materials. Dent Mater 2001;17(4):289295. DOI: 10.1016/s0109-5641(00)00086-5.
6. Graig RG. Restorative dental Materials. The CV Mosby Co: St Louis; 1993.
7. Ban S, Anusavice, KJ. Influence of test method on failure stress of brittle dental materials. J Dent Res, 1990;69:17911799. DOI: 10.1177/00220345900690120201.
8. McLean J. The science and art of dental ceramics. Vol. 1. Quintessence: Chicago; 1979.
9. McLean J. Dental ceramics. Proceedings of the First International Symposium on Ceramics. Quintessence: Chicago; 1983.
10. Mehulić K, Laus-Sosić M. Metal-ceramic bond: how to improve? Minerva Stomatol 2009;58(78):36773. PMID: 19633637.
11. Ashtiani AH, Mardasi N, Fathi A. Effect of multiple firings on the shear bond strength of presintered cobalt-chromium alloy and veneering ceramic. J Prosthet Dent 2021;126(6):803.e1803.e6. DOI: 10.1016/j.prosdent.2021.09.021. Epub 2021 Oct 27. PMID: 34716011.
12. Yilmaz B, Ozçelik TB, Wee AG. Effect of repeated firings on the color of opaque porcelain applied on different dental alloys. J Prosthet Dent 2009;101(6):395404. DOI: 10.1016/S0022-3913(09)60085-8. PMID: 19463667.
13. Ozcelik TB, Yilmaz B, Ozcan I, et al. Colorimetric analysis of opaque porcelain fired to different base metal alloys used in metal ceramic restorations. J Prosthet Dent 2008;99(3):193202. DOI: 10.1016/S0022-3913(08)60043-8.
14. EN ISO 9693. Metal-ceramic dental restorative systems. 2nd edition. International Organization for Standardization: Geneva, Switzerland; 1999.
15. Atluri KR, Vallabhaneni TT, Tadi DP, et al. Comparative evaluation of metal-ceramic bond strengths of nickel chromium and cobalt chromium alloys on repeated castings: An in vitro study. J Int Oral Health 2014;6(5):99103.
16. Beck KA, Sarantopoulos DM, Kawashima I, et al. Elemental release from Co-Cr and Ni-Cr alloys containing palladium. J Prosthet Dent 2012;21:8893.
17. Probster, L, Maiwald, U, Weber H. Three point bending strength of ceramics fused to cast titanium. Eur J Oral Sci 1996;104:313319. DOI: 10.1111/j.1600-0722.1996.tb00083.x.
18. Ghiban A, Moldovan P. Study of corrosion behavior under simulated physiological conditions of dental Co-Cr-Mo-Ti alloys. U.P.B. Sci Bull 2012;74:203214.
19. Vásquez VZ, Ozcan M, Kimpara ET. Evaluation of interface characterization and adhesion of glass ceramics to commercially pure titanium and gold alloy after thermal- and mechanical-loading. Dent Mater 2009;25(2):221231. DOI: 10.1016/j.dental.2008.07.002. Epub 2008 Aug 20.
20. Ohno H, Kanzawa Y. Structural changes in the oxidation zones of gold alloys for porcelain bonding containing small amounts of Fe and Sn. J Dent Res 1985;64(1):6773. DOI: 10.1177/00220345850640011401.
21. Shillingburg HT, Hobo S, Fisher DW. Preparation design and margin distortion in porcelain fused to metal restorations. J Prosthet Dent 1973;29(3):276. DOI: 10.1016/0022-3913(73)90007-3.
22. Brodbelt RHW, O’Brien WJ, Fan RL. Translucency of dental porcelains. J Dental Res 1980;59(1):7075. DOI: 10.1177/00220345800590011101.
23. Asgar K, Arfaei AH. Castability of crown and bridge alloys. J Prosthet Dent1985;54(1):6063. DOI: 10.1016/s0022-3913(85)80071-8.
24. Lee SK, Wilson PR. Fracture of all-ceramic crowns with varying core elastic moduli. Aust Dent J 2000;45(2):103107.
25. Kerber SJ. The complementary nature of X ray photoelectron spectroscopy and angle resolved X ray diffraction. II. Analysis of oxides on dental alloys. J Material Engl Perform 1998;7:334342.
26. Cramber DA, Goldstein RE. Cast - ceramic systems and other alternatives in porcelain and composite inlays and outlays - Esthetic posterior restorations. Quintessence Books: Michigan; 1994.
27. Schweizer DM, Goldstein GR, Ricci JL, et al. Comparison of bond strength of a pressed ceramic fused to metal versus feldspathic porcelain fused to metal. J Posthodont 2009;14(4):239247. DOI: 10.1111/j.1532-849X.2005.00052.x.
28. Akagi M, Okamato Y, Matsura T, et al. Properties of test metal ceramic titanium alloys. J Prosthet Dent 1992;68:462467. DOI: 10.1016/0022-3913(92)90411-3.
29. Rathi S, Parkash H, Chittaranjan B, et al. Oxidation heat treatment affecting metal-ceramic bonding. Indian J Dent Res 2011;22(6):877878. DOI: 10.4103/0970-9290.94664.
30. Li J, Ye X, Li B, et al. Effect of oxidation heat treatment on the bond strength between a ceramic and cast and milled cobalt-chromium alloys. Eur J Oral Sci 2015;123(4):297304. DOI: 10.1111/eos.12199.
31. Patel KA, Mathur S, Upadhyay S. A comparative evaluation of bond strength of feldspathic porcelain to nicel–chromium alloy, when subjected to various surface treatments: An in vitro study. J Indian Prosthod Soc 2015;15(1):5357. DOI: 10.4103/0972-4052.155036.
32. Lahori M, Nagrath R, Sisodia S. The effect of surface treatments on the bond strength of a nonprecious alloy–ceramic interface: An in vitro study. J Indian Prosthod Soc 2014;14(2):151–155. DOI: 10.1007/s13191-013-0285-3.
33. Vojdani M, Shaghaghian S, Khaledi A, et al. The effect of thermal and mechanical cycling on bond strength of a ceramic to nickel - chromium (Ni-Cr) and cobalt-chromium (Co-Cr) alloys. Indian J Dent Res 2012;23(4):509513. DOI: 10.4103/0970-9290.104960.
________________________
© The Author(s). 2023 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by-nc/4.0/), which permits unrestricted use, distribution, and non-commercial reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.