The Journal of Contemporary Dental Practice

Register      Login

SEARCH WITHIN CONTENT

FIND ARTICLE

Volume / Issue

Online First

Archive
Related articles

VOLUME 13 , ISSUE 4 ( July-August, 2012 ) > List of Articles

RESEARCH ARTICLE

Evaluation of the Cadiax Compact® II Accuracy in Recording Preadjusted Condylar Inclinations on Fully Adjustable Articulator

Ahmad Hasan Ahangari, Kianoosh Torabi, Sasan Rasaei Pour, Safoura Ghodsi

Citation Information : Ahangari AH, Torabi K, Pour SR, Ghodsi S. Evaluation of the Cadiax Compact® II Accuracy in Recording Preadjusted Condylar Inclinations on Fully Adjustable Articulator. J Contemp Dent Pract 2012; 13 (4):504-508.

DOI: 10.5005/jp-journals-10024-1176

Published Online: 01-12-2012

Copyright Statement:  Copyright © 2012; The Author(s).


Abstract

Background

Mandibular movement analysis is a critical step in making the functional occlusal morphology and improving the diagnosis and treatment of temporomandibular joint disorders (TMDs). Cadiax Compact® is an electronic condylograph that claims to record the horizontal condylar inclination (HCI), Bennett angle (BA) and relative shape of the articular eminence. This study aims at assessing the accuracy of Cadiax Compact® II in its claimed abilities.

Materials and methods

The electronic condylograph (Cadiax Compact®II) was fitted on the fully adjustable articulator (gamma dental reference- SL). After setting of HCI and BA on the arbitrary degrees, eccentric movements were produced manually on the articulator. The Cadiax recorded these preadjusted angles and the accuracy of its recordings was assessed by comparison of the results with the preadjusted HCI, BA and color inserts as references.

Results

The majority of the comparisons showed statistically significant differences between articulator settings and Cadiax recordings. However, the maximum difference was about 2.5° which seems acceptable for clinical practice.

Conclusion

The obtained results showed that Cadiax Compact is an accurate and reliable instrument for diagnostic purposes, yielding reproducible measurements. Despite this, Cadiax is a technically sensitive device that can preclude its routine usage.

How to cite this article

Ahangari AH, Torabi K, Pour SR, Ghodsi S. Evaluation of the Cadiax Compact® II Accuracy in Recording Preadjusted Condylar Inclinations on Fully Adjustable Articulator. J Contemp Dent Pract 2012;13(4):504-508.


PDF Share
  1. Technical validation of a computerized condylographic system. Int J Stomat Occ Med 2009;1:45-50.
  2. Clinical and instrumental functional analysis for diagnosis and treatment planning. Part 5. Axiography. J Clin Orthod 1988;22:656-67.
  3. Computerized axiography: Principles and methods. Cranio 1991;9:344-55.
  4. Applied Gnathology: Why and how. Dent Clin North Am 1969;13:687-700.
  5. Contemporary fixed prosthodontics (4th ed), Mosby: Elsevier; 2006. Ch 1.
  6. A comparison of protrusive interocclusal records to pantographic tracings. J Prosthet Dent 1989;62:154-56.
  7. The movements of the lower jaw. Boston Med Surg J 1889;121:8-11.
  8. Some graphic records of movements of the mandible in the living subject (and their bearing on the mechanism of the joint and construction of articulators). Dent Cosmos 1905;47:39.
  9. Mandibular movements in three dimensions. J Prosth Dent 1963;13:72-92.
  10. Jaw movements engraved in solid plastic for articular controls. I. Recording apparatus. J Prosthet Dent 1969;22:209-24.
  11. The accuracy and practical value of records of condyle path inclination. J Am Dent Assoc 1949;38:697-710.
  12. Accurate interocclusal records. J Prosthet Dent 1960;10:620-30.
  13. Evaluation of Bennett angle in relation to a sagital path. Osterr Zeitcher F Stomatol 1978;7:8-15.
  14. Clinical evaluation of electronic pantograph in relation to post axes. J Dent Res 1983;62:200.
  15. An in vitro study of an electronic pantograph. J Prosthet Dent 1987;57:577-80.
  16. Clinical evaluation of electronic pantograph with mechanical pantograph. J Dent Res 1983;62:200.
  17. Comparison of an electronic and a mechanical pantograph. Part I: Consistency of an electronic computerized pantograph to record articulator settings. J Prosthet Dent 1986;55(5):570-74.
  18. A comparison of articulator settings obtained by using an electronic pantograph and lateral interocclusal recordings. J Prosthet Dent 1988;60:159-164.
  19. Mechanical and computerized condylography, a comparative study. Inf Orthod Kieferorthop 1995;27:15-23.
  20. Accuracy of recording horizontal condylar inclination and Bennett angle with the Cadiax Compact. J Oral Rehabil 2002;29(11):1076-81.
  21. An in vitro evaluation of the reliability and validity of an electronic pantograph by testing with five different articulators. J Prosthet Dent 2004;92(1):83-89.
  22. Skytting B. Registration of the condyle path inclination: Variations using the Gysi technique. J Prosthet Dent 1960;10:243-47.
  23. A study of the variability of setting a fully adjustable gnathologic articulator to a pantographic tracing. J Prosthet Dent 1977;37(4):460-65.
  24. Errors incurred in programming a fully adjustable articulator with a pantograph. J Prosthet Dent 1986;55(4):427-29.
PDF Share
PDF Share

© Jaypee Brothers Medical Publishers (P) LTD.