The Journal of Contemporary Dental Practice

Register      Login

SEARCH WITHIN CONTENT

FIND ARTICLE

Volume / Issue

Online First

Archive
Related articles

VOLUME 14 , ISSUE 4 ( July-August, 2013 ) > List of Articles

RESEARCH ARTICLE

A Scanning Electron Microscopic Study to evaluate the Efficacy of NaviTip FX in removing the Canal Debris during Root Canal Preparation: An in vitro Study

KG Nischith, Vinay Chandra, Padma Gandi, Anil Kumar Shivanna, Siva Srinivas, S Himgiri

Citation Information : Nischith K, Chandra V, Gandi P, Shivanna AK, Srinivas S, Himgiri S. A Scanning Electron Microscopic Study to evaluate the Efficacy of NaviTip FX in removing the Canal Debris during Root Canal Preparation: An in vitro Study. J Contemp Dent Pract 2013; 14 (4):653-656.

DOI: 10.5005/jp-journals-10024-1380

Published Online: 01-12-2013

Copyright Statement:  Copyright © 2013; The Author(s).


Abstract

Objective

To evaluate the efficacy of NaviTip FX in removing the canal debris during root canal preparation using scanning electron microscopic study

Materials and methods

Thirty single rooted teeth with completely formed apices were used in this study. Standard endodontic access cavity preparations were performed. Then the teeth were randomly divided into two groups: groups 1 and 2 of 15 teeth each group. For group 1, NaviTip FX (brush covered needle) was used to irrigate the canal with 5.25% sodium hypochlorite after each instrument use. For group 2, NaviTip (brushless needle) was used for irrigation following each instrument use. ProTaper rotary files were used for the canal preparation. The teeth were then cleaned and dried before splitting them into two halves. The half with most visible part of the apex was used for scanning electron microscopic evaluation. The results were statistically analyzed using the Mann-Whitney U-test at significance level p < 0.005.

Results

The mean values for coronal and middle third of group 1 showed lower debris scores than group 2 and this difference was statistically significant at a p-value 0.01 and 0.05 respectively, but no significance difference between them at the apical third at a p-value of < 0.05.

Conclusion

The NaviTip FX (brush covered needle) showed effectively better canal wall debris removal than the NaviTip (brushless needle).

How to cite this article

Chandra V, Gandi P, Shivanna AK, Srinivas S, Himgiri S, Nischith KG. A Scanning Electron Microscopic Study to evaluate the Efficacy of NaviTip FX in removing the Canal Debris during Root Canal Preparation: An in vitro Study. J Contemp Dent Pract 2013;14(4):653-656.


PDF Share
  1. Apical limit of root canal instrumentation and obturation, Part 2. A histological study. Int Endod J 1998;31:394-409.
  2. Aprimary observation on the preparation and obturation of oval canals. Int Endod J 2001;34:137-141.
  3. Effectiveness in cleaning oval-shaped root canals using Anatomic Endodontic Technology, Profile and manual instrumentation: a scanning electron microscopic study. Int Endod J 2005;38:356-363.
  4. A scanning electron microscopic evaluation of four root canal irrigation regimens. J Endod 1987;13:147-157.
  5. A scanning electron microscopic comparison of a high volume final flush with several irrigating solutions: Part 3. J Endod 1983;9:137-142.
  6. Root canal cleanliness after preparation with different endodontic hand pieces and hand instruments: a comparative SEM investigation. J Endod 1997;23:301-306.
  7. A comparative scanning electron microscopic investigation of the efficacy of manual and automated instrumentation of root canals. J Endod 2000;26:660-664.
  8. In vivo debridement efficacy of ultrasonic irrigation following hand rotary instrumentation in human mandibular molars. J Endod 2005;31:166-170.
  9. Scanning electron microscopic study of the efficacy of various irrigating solutions. J Endod 1975;1:127-135.
  10. The solvent action of sodium hypochlorite on pulp tissue of extracted teeth. Oral Surg Oral Med Oral Pathol 1997;31:96-103.
  11. Chemomechanical removal of pulpal and dentinal debris with sodium hypochlorite and hydrogen peroxide vs normal saline solution. J Endod 1997;3:49-53.
  12. Evaluation of canal cleanliness and smear layer removal after the use of the Quantec—E irrigation system and syringe: a comparative scanning electron microscope study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2003;96:614-617.
  13. The use of a cotton-wrapped file to evacuate necrotic pulp tissue. Oral Surg 1979;48:77-78.
  14. Effectiveness of a brush in removing post instrumentation canal debris. J Endod 1990;16:323-327.
  15. Comparison of root canal preparation using Race and ProTaper rotary NiTi instruments. Int Endod J 2005;38:8-16.
  16. A preliminary scanning electron microscopic study of root canals after endodontic procedures. J Endod 1975;1:238-242.
  17. Microscopic and macroscopic investigation on results of mechanical preparation of root canals. Oral Surg Oral Med Oral Pathol 1968;25:108-116.
  18. Microscopic measurements of root canal preparations following instrumentation. J Brit Endo Soc 1968;2:41-49.
  19. Determination of the minimum instrumentation size for penetration of irrigants to the apical third of root canal systems. J Endod 2006;32:417-420.
  20. The effectiveness of increased apical enlargement in reducing intracanal bacteria. J Endod 2002;28:779-783.
  21. Influence of instrument size on root canal debridement. J Endod 2004;30:110-112.
  22. Efficacy of bacterial removal from instrumented root canals in vitro related to instrumentation technique and size. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2002;94:366-371.
PDF Share