VOLUME 19 , ISSUE 7 ( 2018 ) > List of Articles

ORIGINAL RESEARCH

Sterilization and Biologic Monitoring in Private Dental Clinics in Lebanon

Zeina Majzoub, Jihad Dagher, Charles Sfeir, Ahmad Abdallah

Keywords : Biologic indicator, Dentists, Lebanon, Private dental clinic, Sterilization

Citation Information : Majzoub Z, Dagher J, Sfeir C, Abdallah A. Sterilization and Biologic Monitoring in Private Dental Clinics in Lebanon. J Contemp Dent Pract 2018; 19 (7):853-861.

DOI: 10.5005/jp-journals-10024-2347

License: CC BY-NC 3.0

Published Online: 01-03-2018

Copyright Statement:  Copyright © 2018; The Author(s).


Abstract

Aim: The aim of the present study was to evaluate sterilization practices and effectiveness in the Lebanese private dental sector and identify potential factors contributing to sterilization failure. Materials and methods: A 13-item questionnaire consisting of four demographic/professional questions and nine questions related to sterilization practices along with self-contained biologic indicators (SCBIs) were delivered to a representative sample of Lebanese private offices. Univariate statistics and bivariate analyses were performed to compare sterilization failure rates according to demographic, professional, and sterilization-related conditions. Results: Out of the 560 dentists contacted, 205 dentists returned the completed questionnaires and undamaged processed SCBIs. The tested autoclaves (n = 134) were mostly dynamic air removal (69.4%) and had a mean age of 10.5 ± 6.9 years. The dry heat ovens (n = 71) were all static air and had 12.9 ± 8.1 years. The dental assistants performed the routine sterilization procedures in nearly 62% of the practices and sterilization cycles were run 4 to 6 times per week in 75% of the offices. Correct temperature/time ratios were applied in 97% of the autoclaves and 80.3% of the ovens. Few dental practices reported having preventive maintenance (17.9% for the autoclaves and 14.1% for the ovens). Routine monitoring of sterilizer efficacy was infrequently performed and was mostly conducted using physical indicators. Sterilization failure rate was higher for the ovens (16.9%) than for the autoclaves (7.5%). Incorrect temperature/time ratio was the main significant factor associated with sterilization failures. Conclusion: The present study demonstrated a relatively high rate of sterilization failures in the Lebanese private dental sector and identified the human error in setting sterilization cycle parameters as the predominant cause of failure. These findings should prompt actions toward increasing knowledge of the sterilization processes and their monitoring among dental professionals and improving the quality control of sterilization through collaborative efforts among health authorities, dental schools, and associations. Clinical significance: This study presents the first published data relative to sterilization practices and effectiveness in private Lebanese dental offices and provides a rationale to implement biologic monitoring protocols in Lebanon as long practiced in developed countries.


PDF Share
  1. Rutala, WA.; Weber, DJ.; the Healthcare Infection Control Practices Advisory Committee (HICPAC). Guideline for disinfection and sterilization in healthcare facilities. Clifton Road Atlanta (GA): CDC; 2008. p. 58-79. [cited 2017 Feb 15]. Available from: https://www.cdc.gov/infectioncontrol/guidelines/disinfection/.
  2. Association for the Advancement of Medical Instrumentation. ANSI/AAMI:ST79/2017: comprehensive guide to steam sterilization and sterility assurance in health care facilities. Arlington (VA): AAMI; 2017. p. 79-101. Available from: https://standards.aami.org/higherlogic/ws/.../Public%20 review%20CDV-2%20ST79.pdf
  3. Schneider PM, Reich RR, Kirckof SS, Foltz WG. Performance of various steam sterilization indicators under optimum and sub-optimum exposure conditions. Am J Infect Control 2005 Jun;33(5 Suppl 2):S55-S67.
  4. Sella SR, Vandenberghe LP, Soccol CR. Bacillus atrophaeus: main characteristics and biotechnological applications—a review. Crit Rev Biotechnol 2015 Oct;35(4);533-545.
  5. Scheutz F, Reinholdt J. Outcome of sterilization by steam autoclaves in Danish dental offices. Scand J Dent Res 1988 Apr;96(2):167-170.
  6. Skaug N. Proper monitoring of sterilization procedures used in oral surgery. Int J Oral Surg 1983 Jun;12(3):153-158.
  7. Acosta-Gío AE, Mata-Portuguez VH, Herrero-Farías A, Sánchez Pérez L. Biologic monitoring of dental office sterilizers in Mexico. Am J Infect Control 2002 May;30(3):153-157.
  8. Kohn WG, Collins AS, Cleveland JL, Harte JA, Eklund KJ, Malvitz DM. Guidelines for infection control in dental health-care settings—2003. MMWR Recomm Rep 2003 Dec;52(RR17):1-61.
  9. de Nardo AP, Romano TG, Dias AGA, Guimarães G. Sterilization in the dental private sector. RGO—Rev Gaúcha Odontol 2013 Jan-Mar;61(1):47-53.
  10. Dagher J, Sfeir C, Abdallah A, Majzoub Z. Infection control measures in private dental clinics in Lebanon. Int J Dent 2017 May;2017:5057248.
  11. Simonsen RJ, Schachtele CF, Joos RW. An evaluation of sterilization by autoclave in dental offices. J Dent Res 1979 Jan;58:400.
  12. Palenik CJ, King TN, Newton CW, Miller CH, Koerber LG. A survey of sterilization practices in selected endodontic offices. J Endod 1986 Jan;12(5):206-209.
  13. Nickerson A, Bhuta P, Orton G, Alvin B. Monitoring dental sterilizers’ effectiveness using biological indicators. J Dent Hyg 1990 Feb;64(2):69-73.
  14. Hastreiter RJ, Molinari JA, Falken MC, Roesch MH, Gleason MJ, Merchant VA. Effectiveness of dental office instrument sterilization procedures. J Am Dent Assoc 1991 Oct;122(10):51-56.
  15. Molinari JA, Gleason MJ, Merchant VA. Sixteen years of experience with sterilization monitoring. Compendium 1994 Dec;15(12):1422-1424.
  16. Hackney RW Jr, Crawford JJ, Tulis JJ. Using a biological indicator to detect potential sources of cross-contamination in the dental operatory. J Am Dent Assoc 1998 Nov;129(11):1567-1577.
  17. Cuny E. The use of a process challenge device in dental office gravity displacement tabletop sterilizers. Am J Infect Control 2015 Oct;43(10):1131-1133.
  18. McErlane B, Rosebush WJ, Waterfield JD. Assessment of the effectiveness of dental sterilizers using biological monitors. J Can Dent Assoc 1992 Jun;58(6):481-483.
  19. Skaug N, Lingaas E, Nielsen O, Palenik CJ. Biological monitoring of sterilizers and sterilization failures in Norwegian dental offices in 1985 and 1996. Acta Odontol Scand 1999 Aug;57(4):175-180.
  20. Andrés MT, Tejerina JM, Fierro JF. Reliability of biologic indicators in a mail-return sterilization-monitoring service: a review of 3 years. Quintessence Int 1995 Dec;26(12):865-870.
  21. Werner HP, Kindt R, Borneff J. Testing the sterilisation effect of autoclaves by means of biological indicators (author's transl). Zentralbl Bakteriol Orig B 1975 Jul;160(4-5):458-472.
  22. de Graaff J, Eijkman MA. Quality of the sterilization process in the dental practice. Ned Tijdschr Tandheelkd 1997 Sep;104(9):342-345.
  23. Tavares Sdo S, de Sousa JT, Tipple AF, e Souza AC, Pimenta FC, Anders PS. Efficacy of the pasteur oven as sterilization equipment in dental offices. Rev Esc Enferm USP 2008 Mar;42(1): 160-167.
  24. Podgórska M, Jakimiak B, Röhm-Rodowald E, Chojecka A. Assessment of disinfection and sterilization processes in dental practice as important factors in prevention of infections. Przegl Epidemiol 2009 Jan;63(4):545-550.
  25. Röhm-Rodowald E, Jakimiak B, Chojecka A, Zmuda- Baranowska M, Kanclerski K. Assessment of decontamination processes: cleaning, disinfection and sterilization in dental practice in Poland in the years 2011-2012. Przegl Epidemiol 2012 Jan;66(4):635-641.
  26. Patiño-Marín N, Martínez-Castañón GA, Zavala-Alonso NV, Medina-Solís CE, Torres-Méndez F, Cepeda-Argüelles O. Biologic monitoring and causes of failure in cycles of sterilization in dental care offices in Mexico. Am J Infect Control 2015 Oct;43(10):1092-1095.
  27. Corrêa EG, de Castilho AR, Pereira CV. Indicadores químicos e biológicos da eficácia de esterilização por autoclave ou estufa. Rev Odontol Ciênc 2009 Apr;24(2):156-160.
  28. Prado ME, Santos SS. Avaliação das condições de esterilização de materiais odontológicos, em consultórios na cidade de Taubaté. Rev Bioc Taubaté 2002 Aug;8(1):61-70.
  29. McCauley K. A review of the newly revised ISO 11138 series on the manufacture of biological indicators. SporeNews Biol Indic Newslett MesaLabs 2017;13(2):1-3.
  30. Alfa MJ, Olson N, DeGagne P, Jackson M. Evaluation of rapid readout biological indicators for 132 degrees C gravity and 132 degrees C vacuum-assisted steam sterilization cycles using a new automated fluorescent reader. Infect Control Hosp Epidemiol 2002 Jul;23(7):388-392.
PDF Share
PDF Share

© Jaypee Brothers Medical Publishers (P) LTD.