VOLUME 20 , ISSUE 7 ( July, 2019 ) > List of Articles

CLINICAL TECHNIQUE

A Consolidated Pulp Test System Including Flowmetry, Pulse Oximetry, and Thermometry

Hamid Jafarzadeh, Farzaneh Iusefipour, Mohammad-Ebrahim Zirouhi, Jun-Ichiro Kinoshita, Atsufumi Manabe, Mikihiro Kobayashi, Amir Azarpazhooh

Keywords : Endodontics, Flowmetry, Pulp test, Pulp vitality, Pulse oximetry, Thermometry

Citation Information : Jafarzadeh H, Iusefipour F, Zirouhi M, Kinoshita J, Manabe A, Kobayashi M, Azarpazhooh A. A Consolidated Pulp Test System Including Flowmetry, Pulse Oximetry, and Thermometry. J Contemp Dent Pract 2019; 20 (7):873-877.

DOI: 10.5005/jp-journals-10024-2614

License: CC BY-NC 4.0

Published Online: 01-08-2019

Copyright Statement:  Copyright © 2019; The Author(s).


Abstract

Aim: The aim of this study is to report manufacturing a consolidated pulp test system that includes thermometry, flowmetry, and pulse oximetry. Background: The ideal method for the evaluation of pulp vitality should be objective, noninvasive, easy to use, reliable, and painless. Currently, the most commonly used tests (such as electricity, heat, and cold tests) depend on patients’ sensibility and are highly subjective. They only measure a neural response and do not indicate the actual biologic status of the pulp. It has, therefore, been suggested that vitality tests such as flowmetry, pulse oximetry, thermometry, and photoplethysmography be used. Some research has been done on these systems; however, their high costs and space need to maintain all of them that have been obstacles to their use. Technique: This report describes designing and manufacturing a novel system for evaluating pulp vitality involving the use of three methods (flowmetry, pulse oximetry, and thermometry) combined in a single small system using only two probes. Conclusion: The consolidated pulp tests system may be accurate in determining the pulp vitality. However, after clinical use, some changes may be necessary for improvement of the system.


HTML PDF Share
  1. Jafarzadeh H, Abbott PV. Review of pulp sensibility tests. Part I: general information and thermal tests. Int Endod J 2010;43:738–762. DOI: 10.1111/j.1365-2591.2010.01754.x.
  2. Jafarzadeh H, Abbott PV. Review of pulp sensibility tests. Part II: electric pulp tests and test cavities. Int Endod J 2010;43:945–958. DOI: 10.1111/j.1365-2591.2010.01760.x.
  3. Jafarzadeh H, Rosenberg PA. Pulse oximetry: review of a potential aid in endodontic diagnosis. J Endod 2009;35:329–333. DOI: 10.1016/j.joen.2008.12.006.
  4. Jafarzadeh H. Laser Doppler flowmetry in endodontics: a review. Int Endod J 2009;42:476–490. DOI: 10.1111/j.1365-2591.2009.01548.x.
  5. Jafarzadeh H, Udoye CI, et al. The application of tooth temperature measurement in endodontic diagnosis: a review. J Endod 2008;34:1435–1440. DOI: 10.1016/j.joen.2008.09.011.
  6. Miwa Z, Ikawa M, et al. Pulpal blood flow in vital and nonvital young permanent teeth measured by transmitted-light photoplethysmography: a pilot study. Pediatr Dent 2002;24:594–598.
  7. McMorrow RC, Mythen MG. Pulse oximetry. Curr Opin Crit Care 2006;12:269–271. DOI: 10.1097/01.ccx.0000224873.16700.78.
  8. Noblett WC, Wilcox LR, et al. Detection of pulpal circulation in vitro by pulse oximetry. J Endod 1996;22:1–5. DOI: 10.1016/S0099-2399(96)80226-3.
  9. Dastmalchi N, Jafarzadeh H, et al. Comparison of the Efficacy of a Custom-made Pulse Oximeter Probe with Digital Electric Pulp Tester, Cold Spray, and Rubber Cup for Assessing Pulp Vitality. J Endod 2012;38:1182–1186. DOI: 10.1016/j.joen.2012.06.012.
  10. Yeh Y, Cummins HZ. Localized fluid flow measurements with an He-Ne laser spectrometer. Appl Phys Lett 1964;4:176–178. DOI: 10.1063/1.1753925.
  11. Gazelius B, Olgart L, et al. Non-invasive recording of blood flow in human dental pulp. Endod Dent Traumatol 1986;2:219–221. DOI: 10.1111/j.1600-9657.1986.tb00148.x.
  12. Gazelius B, Olgart L, et al. Restored vitality in luxated teeth assessed by laser Doppler flowmeter. Endod Dent Traumatol 1988;4:265–268. DOI: 10.1111/j.1600-9657.1988.tb00645.x.
  13. Olgart LM, Edwall B, et al. Neurogenic mediators in control of pulpal blood flow. J Endod 1989;15:409–412. DOI: 10.1016/S0099-2399(89)80173-6.
  14. Ebihara A, Tokita Y, et al. Pulpal blood flow assessed by laser Doppler flowmetry in a tooth with a horizontal root fracture. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1996;81:229–233. DOI: 10.1016/S1079-2104(96)80420-3.
  15. Yanpiset K, Vongsavan N, et al. Efficacy of laser Doppler flowmetry for the diagnosis of revascularization of reimplanted immature dog teeth. Dent Traumatol 2001;17:63–70. DOI: 10.1034/j.1600-9657.2001.017002063.x.
  16. Howell RM, Duell RC, et al. The determination of pulp vitality by thermographic means using cholesteric liquid crystals. A preliminary study. Oral Surg Oral Med Oral Pathol 1970;29:763–768. DOI: 10.1016/0030-4220(70)90275-6.
  17. Stoops LC, Scott Jr D. Measurement of tooth temperature as a means of determining pulp vitality. J Endod 1976;2:141–145. DOI: 10.1016/S0099-2399(76)80011-8.
  18. Banes JD, Hammond HL. Surface temperatures of vital and nonvital teeth in humans. J Endod 1978;4:106–109. DOI: 10.1016/S0099-2399(78)80199-X.
  19. Fanibunda KB. A laboratory study to investigate the differentiation of pulp vitality in human teeth by temperature measurement. J Dent 1985;13:295–303. DOI: 10.1016/0300-5712(85)90024-7.
  20. Fanibunda KB. The feasibility of temperature measurement as a diagnostic procedure in human teeth. J Dent 1986;14:126–129. DOI: 10.1016/0300-5712(86)90077-1.
  21. Fanibunda KB. Diagnosis of tooth vitality by crown surface temperature measurement: a clinical evaluation. J Dent 1986;14:160–164. DOI: 10.1016/0300-5712(86)90018-7.
  22. Nafisi V, Sakhavi S. Designing equipment for monitoring and analyzing the pulse signal based on the Iranian traditional medicine. 4th Iranian Conference on eHealth and ICT Application in Medical Sciences, Mashhad, 2011.
  23. Nafisi V. Designing and manufacturing pulse-meter for using in traditional medicine. Iranian Research Organization for Science and Technology 2013.
  24. LPC1768 user manual, available online from: http://www.nxp.com/documents/user_manual/UM10360.pdf.
  25. Yousefipoor F, Mirhadi M, et al. Device Design and Manufacturing for Measurement of Kinematic Specification of Hand Movement. Majlesi Journal of Multimedia Processing 2013;4(2):1–6.
  26. Ash M, Nelson SJ. Wheeler's Dental Anatomy. Physiology, and Occlusion, 9th ed. St. Louis: Saunders, 2010.
  27. Evans D, Reid J, et al. A comparison of laser Doppler flowmetry with other methods of assessing the vitality of traumatised anterior teeth. Endod Dent Traumatol 1999;15:284–290. DOI: 10.1111/j.1600-9657.1999.tb00789.x.
  28. Crandell CE, Hill RP. Thermography in dentistry: a pilot study. Oral Surg Oral Med Oral Pathol 1966;21:316–320. DOI: 10.1016/0030-4220(66)90064-8.
  29. Brown AC, Goldberg MP. Surface temperature and temperature gradients of human teeth in situ. Arch Oral Biol 1966;11:973–982. DOI: 10.1016/0003-9969(66)90199-3.
  30. Hartley JL, Stanfill DF, et al. Thermography of the human dentition. SAM-TR-67-57, USAF School of Aerospace Medicine 1967;1–40.
  31. Stanfill DF, Plakun BD. Feasibility study, thermography of human dentition. Barnes Engineering Co., Project 3812. Air Force Project Task no. 799601. 1966.
  32. Herrmann M. Temperaturverhaltnisse an der Mundschleimhaut der Zunge und er Zahnen. Dtsch Zahnaerztl Z 1953;8:539–543.
  33. Goldberg M, Brown AC. Human tooth surface temperature. Physiologist 1965;8:175.
  34. Baumann G. Temperaturmessungen an vitalen und devitalen Zahnen. Inaugural dissertation. Mainz: Johannes Gutenberg Universitat, 1952.
  35. Stoops LC, Scott Jr D. Measurement of tooth temperature as a means of determining pulp vitality. J Endod 1976;2:141–145. DOI: 10.1016/S0099-2399(76)80011-8.
  36. Banes JD, Hammond HL. Surface temperatures of vital and nonvital teeth in humans. J Endod 1978;4:106–109. DOI: 10.1016/S0099-2399(78)80199-X.
  37. Howell RM, Duell RC, et al. The determination of pulp vitality by thermographic means using cholesteric liquid crystals. A preliminary study. Oral Surg Oral Med Oral Pathol 1970;29:763–768. DOI: 10.1016/0030-4220(70)90275-6.
  38. Kahan RS, Gulabivala K, et al. Evaluation of a pulse oximeter and customized probe for pulp vitality testing. J Endod 1996;22:105–109. DOI: 10.1016/S0099-2399(96)80283-4.
  39. Goho C. Pulse oximetry evaluation of vitality in primary and immature permanent teeth. Pediatr Dent 1999;21:125–127.
  40. Gopikrishna V, Tinagupta K, et al. Comparison of electrical, thermal, and pulse oximetry methods for assessing pulp vitality in recently traumatized teeth. J Endod 2007;33:531–535. DOI: 10.1016/j.joen.2007.01.014.
  41. Gopikrishna V, Tinagupta K, et al. Evaluation of efficacy of a new custom-made pulse oximeter dental probe in comparison with the electrical and thermal tests for assessing pulp vitality. J Endod 2007;33:411–414. DOI: 10.1016/j.joen.2006.12.003.
PDF Share
PDF Share

© Jaypee Brothers Medical Publishers (P) LTD.