The Journal of Contemporary Dental Practice

Register      Login

SEARCH WITHIN CONTENT

FIND ARTICLE

Volume / Issue

Online First

Archive
Related articles

VOLUME 14 , ISSUE 5 ( September-October, 2013 ) > List of Articles

CASE REPORT

Removal of a Fractured Locator Abutment Screw Fragment

Ilser Turkyilmaz, Matthew Joseph Vierra, Neset Volkan Asar

Citation Information : Turkyilmaz I, Vierra MJ, Asar NV. Removal of a Fractured Locator Abutment Screw Fragment. J Contemp Dent Pract 2013; 14 (5):968-972.

DOI: 10.5005/jp-journals-10024-1435

Published Online: 01-04-2014

Copyright Statement:  Copyright © 2013; The Author(s).


Abstract

Background

Although, dental implants have been a predictable option to support several types of restorations, mechanical problems are not uncommon. Retaining and abutment screw loosenings/fractures are one of the most common technical problems.

Aim

To present a step-by-step technique in order to remove a broken locater abutment screw fragment.

Materials and methods

A 56-year-old, edentulous man with a fractured locator abutment screw in an implant was referred to our clinic from a private practice. His dentist placed a narrow platform locator abutment onto a regular platform implant and the fracture occurred. Attempts to remove the fractured fragment using a periodontal probe and a manual screw removal driver were unsuccessful.

Results

This clinical report presents a situation in which a fractured locater abutment screw fragment was successfully removed using an implant repair kit.

Conclusion

A fractured locater abutment screw fragment was successfully removed using an implant repair kit including drills, drill guides and tapping instruments. To avoid similar complications, it is recommended to use appropriate manufacturer specified torque values in conjunction with compatible restorative components.

Clinical significance

Retaining and abutment screw loosenings/fractures are one of the most common mechanical problems associated with the implant components. This clinical report showed how to remove a broken screw fragment using proper armamentarium and technique, which might help clinicians eliminate similar problems.

How to cite this article

Turkyilmaz I, Vierra MJ, Asar NV. Removal of a Fractured Locater Abutment Screw Fragment. J Contemp Dent Pract 2013;14(5):968-972.


PDF Share
  1. A 5-year prospective multicenter study on 1-stage smooth-surface Branemark system implants with early loading in edentulous mandibles. Int J Oral Maxillofac Implants 2008;23:481-486.
  2. Long-term success of 6 implants supporting a mandibular screw-retained fixed dental prosthesis: a clinical report. J Prosthet Dent 2012;107:280-283.
  3. A 4-year prospective clinical and radiological study of maxillary dental implants supporting single-tooth crowns using early and delayed loading protocols. Clin Implant Dent Relat Res 2007;9:222-227.
  4. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res 2008;19:119-130.
  5. Biomechanical implant treatment complications: a systematic review of clinical studies of implants with at least 1 year of functional loading. Int J Oral Maxillofac Implants 2012;27:894-904.
  6. Clinical complications with implants and implant prostheses. J Prosthet Dent 2003;90:121-132.
  7. Removal of a fractured implant screw: a clinical report. J Prosthet Dent 2004;91:513-517.
  8. The effect of implant diameter, restoration design, and occlusal table variations on screw loosening of posterior single-tooth implant restorations. J Oral Implantol 2001;27:63-72.
  9. Analysis of incidence and associated factors with fractured implants: a retrospective study. Int J Oral Maxillofac Implants 2000;15:662-667.
  10. Evaluation of stress distribution in overdenture-retaining bar with different levels of vertical misfit. J Prosthodont 2011;20:280-285.
  11. Mechanical complications of dental implants. Clin Oral Implants Res 2000;11:156-158.
  12. Load fatigue performance of four implant-abutment interface designs: effect of torque level and implant system. Int J Oral Maxillofac Implants 2008;23:253-262.
  13. Load fatigue performance of a single-tooth implant abutment system: effect of diameter. Int J Oral Maxillofac Implants 2006;21:929-936.
  14. Retrieval of a fractured abutment screw thread from an implant: a case report. Br Dent J 2008;204:177-180.
  15. Retrieval technique for fractured implant screws. J Prosthet Dent 2001;86:549-550.
  16. Fractured prosthetic abutments in osseointegrated implants: a technical complication to cope with. Clin Oral Implants Res 2000;11:163-170.
  17. Elongation and preload stress in dental and abutment screws. Int J Oral Maxillofac Implants 1995;10:529-536.
  18. Torque generated by handheld screwdrivers and mechanical torquing devices for osseointegrated implants. Int J Oral Maxillofac Implants 1994;9:149-155.
PDF Share
PDF Share

© Jaypee Brothers Medical Publishers (P) LTD.