VOLUME 14 , ISSUE 6 ( November-December, 2013 ) > List of Articles

RESEARCH ARTICLE

In vitro Retention Loss of Attachment-retained Removable Partial Denture

Gülsen Can, Baransel Özmumcu, Pinar Altinci

Citation Information : Can G, Özmumcu B, Altinci P. In vitro Retention Loss of Attachment-retained Removable Partial Denture. J Contemp Dent Pract 2013; 14 (6):1049-1053.

DOI: 10.5005/jp-journals-10024-1449

Published Online: 01-04-2014

Copyright Statement:  Copyright © 2013; The Author(s).


Abstract

Aim

To investigate the importance of attachment types on the retention loss of extracoronal attachment-retained removable partial dentures depending on the usage period.

Materials and methods

In order to observe the retention loss of 5 different attachments (OT Strategy, OT Strategy-metal protected, Vario-stud-snap and Vario-soft 3 and ERA-RV) over time, attachment-retained partial dentures representing Kennedy II mod. I case were placed in a custom-made, retention test machine. For each minute, eight separating and joining movements were performed and retention values (Newton) of the attachments were recorded by computer. The retention tests implemented in 540,1080 and 2160 cycles. The data were evaluated statistically according to the two-way ANOVA and Tukey parametrical tests.

Results

The slide type attachment providing the best retention force was observed to be the most worn out by this process (p < 0.01) while the ball type attachments, which typically have the lesser retention force, showed less retention loss (p < 0.01).

Conclusion

It can be concluded that the retention attributes of the attachment-retained dentures were affected by the specific type of precision attachment as well as the usage period.

Clinical significance

Precision attachments with ball-type plastic matrices may be recommended for the clinical use due to their retention stability over time.

How to cite this article

Can G, Özmumcu B, Altinci P. In vitro Retention Loss of Attachment-retained Removable Partial Denture. J Contemp Dent Pract 2013;14(6):1049-1053.


PDF Share
  1. Direct Retainers, indirect retainers and tooth replacements. In: Stewart's clinical removable partial prosthodontics. 3rd ed. Chicago: Quintessence; 2003. p.19.
  2. Direct Retainers. In: Mc Cracken's removable partial prosthodontics. 11th ed. Missouri: Mosby. 2005.p. 79-81.
  3. A classification of precision attachments. J Prosthet Dent 1987;58:322-327.
  4. Treatment planning. In: Precision attachments: a link to successful restorative treatment. London: Quintessence; 1999.p. 11.
  5. Section I. Basic science for dental materials. In: Introduction to dental materials. 2nd ed. London: Mosby; 2002, p. 43-44.
  6. Thermoplastic materials. In: Plastics. Material and processing. New Jersey: Pearson Education; 2006.p.265-268.
  7. Load transfer by a maxillary distal-extension removable partial denture with cap and ring extracoronal attachments. J Prosthet Dent 1992;68:784-789.
  8. Wear behavior of precision attachments. Int J Prosthodont 1999;12:409-412.
  9. Extracoronal resilient attachments in distal-extension removable partial dentures. Quintessence Int 2000;31:311-317.
  10. Retention guide for resilient dental attachments. J Prosthet Dent 2004;92:93-94.
  11. Effect of splinting on load distribution of extracoronal attachments with distal extension prosthesis in vitro. J Prosthet Dent 1996;76:315-320.
  12. Influence of attachment wear on retention of mandibular overdenture. J Oral Rehabil 2007;34:41-51.
  13. Retention and wear of precision-type attachments. J Prosthet Dent 1983;49:28-34.
  14. Artificial saliva substitutes and mineral dissolution. J Oral Rehabil 2001;28:728-731.
  15. In vitro wear of different material combinations of intracoronal precision attachments. Int J Prosthodont 2006;19:330-332.
  16. Retention of ERA direct overdenture attachments before and after fatigue loading. Int J Prosthodont 1997;10:123-130.
  17. Effect of friction coefficient on Akers clasp retention. J Prosthet Dent 1997;78:22-27.
  18. Analysis of potential lubricants for in vitro wear testing. Dent Mater 2006;22:77-83.
  19. Effect of a resilient layer in a removable partial denture base on stress distribution to the mandible. J Prosthet Dent 1980;44:17-20.
  20. In vitro force changes of prefabricated attachments for overdentures. J Oral Rehabil 2003;30:671-678.
  21. Retention profile of stud-type precision attachments. Int J Prosthodont 1991;4:175-179.
  22. Precision attachments retained removable partial dentures: Part 1. Technical long-term study. Int J Prosthodont 1991;4:249-257.
  23. Precision attachments retained removable partial dentures: Part 2. Long-term study of ball attachments. Int J Prosthodont 1995;8:21-28.
  24. Precision attachments retained removable partial dentures: Part 3. General practitioner results up to 2 years. Int J Prosthodont 1998;11:574-579.
  25. Comparative study of two precision overdentures attachment design. J Prosthet Dent 1996;76:145-152.
  26. Load transfer by a maxillary distal-extension removable partial denture with cap and ring extracoronal attachments. J Prosthet Dent 1992;68:784-789.